Stotles logo
Stale pre-tender

Bespoke MSCIII cabinet

Published

Value

50,000 GBP

Description

UKHSA Biosafety Group at Porton Down are looking for a bespoke class III safety cabinet. This equipment will be used for a range of activities including aerobiology and the test and validation of equipment and materials for commercial, research and incident response work. Whilst some of the work will be with pathogens and require a high level of containment, some of it will be using non-pathogenic agents, using the cabinet as a clean enclosure. This will include operation of the cabinet under conditions not normally used at CL3, such as working within the cabinet without the ventilation running. Lot 1: Due to the specific needs and requirements, the solution must comprise the following specifications:<br/>• Internal dimensions of 1600mm wide, 760mm deep and 820mm high.<br/>• Single, non-ventilated pass box. <br/>• Removable front window - being easily removed and replaced.<br/>• Window will have four 200 mm diameter glove ports (with a double o-ring attachment) at a standard height and 2 or 3 fitted in the top half of the window (position and number of ports will be determined and confirmed via simple mock-up).<br/>• Manufactured and tested in accordance to EN 12469:2000.<br/>• The air change rate within the cabinet must not be less than 6 meters cubed per minute for each cubic metre of cabinet volume.<br/>• The working pressure in the cabinet must not be less than -200 Pa (with respect to the laboratory). A manometer capable of measuring from -500 Pa to +500 Pa must be mounted on the cabinet to give a visual indication of the pressure in the interior.<br/>• The cabinet must be supplied with HEPA (High-efficiency particulate air) filtration to Class H-14 99,995% MPPS overall efficiency. Integrity testing of filters using Dispersed Oil Particulate (DOP) scanning of the downstream side of filters, seals, gaskets and construction joints must ensure that no leak penetrations produced are greater than 0.003% of the upstream challenge aerosol. <br/>• Cabinet must be fitted with double HEPA on exhaust. <br/>• Both HEPAs must be independently testable and must be mounted in accordance to section 7.5 of BS EN 12469:2000. <br/>• The cabinet must be fitted with injection and extraction ports for use with an external hydrogen peroxide fumigation systems (1¼” Camlock fittings). Injection to be through the inlet filter and removal post installed extract filters. <br/>• Connection to allow for monitoring of pressure.<br/>• The cabinet must be supplied with clear, easy to operate control panels, visible from the operators seated position and easy to access.<br/>• The main chamber will be fitted with 2 x 1¼” Camlock fittings and 2 compressed gas fittings with solenoid valve linked to alarms.<br/>• The main chamber will have an accessible, removable service plate 40cm x 20cm to allow for flexibility in the future to add additional services, initially provided with the following fittings; 2 compressed air fitting, 4 x 1 1/2” sanitary fitting, 4 x 8mm hose bar fittings, three USB connections, allowing space for future additional fittings as required. <br/>• Fasteners must be stainless steel grade (316). All internal surfaces and fittings should be easily cleanable, and suitably finished to minimise the possibility of damage to the cabinet gauntlets and to facilitate effective cleaning.<br/>• The cabinet must be fitted with 4 min no internal electrical sockets with the following requirements: <br/>- independently switchable; <br/>- 13 amp fused supply; <br/>- 1x Fumigation socket;<br/>- Installed in suitable positions to allow easy access during work (usually on the side panels);<br/>- Splash proof but sockets with fully sealed covers should be avoided;<br/>- Socket covers must allow for small equipment charging devices to be plugged in directly and accommodate a variety of sizes.<br/>• The cabinet must be resistant to disinfectants, including the following chemicals: Sodium hypochlorite 10% solution, Isopropanol, industrial methylated spirit (IMS), Formaldehyde, Hydrogen Peroxide, quaternary ammonias and other disinfectants. The cabinet must be able to withstand repeated fumigation using Formaldehyde or Hydrogen Peroxide systems. <br/>• Performance factory quality assurance documentation (e.g final release record form) must be provided, this is to include but not limited to evidence of HEPA integrity, airflows, pressure decay test, restart automatically after power failure. <br/>• The cabinet will be connected to local ventilation systems by means of an exhaust thimble.<br/>• After installation the cabinet must retain all the exhaust air during fumigation degas. This must be confirmed at the commissioning stage.<br/>• The cabinet must be fitted with shut-tight damper (100% gas tight) on both supply and extract for use during fumigation.<br/>• The cabinet must have the option to be bench mounted or stand mounted.<br/>• The cabinet must be subjected to a pressure hold test at commissioning and then every service. This must be carried out at a positive pressure of >250 Pa, and the cabinet must hold pressure to within +/-10% of the test pressure for not less than 30 minutes. DOP scan test must be carried out at a positive pressure of 250 Pa. All joints, seals and penetrations must be scanned as part of the installation at 250 Pa positive pressure. The recorded penetration around all joints, seals and penetrations must be less than 0.003% of the challenge aerosol. <br/>• Cables are not permitted to pass through the containment barrier, all electrical connections though the cabinet walls must be made with fully sealed bulkhead connections. <br/>• The cabinet is expected to run continuously between service intervals for at least 12 months without need for maintenance.<br/>• There will be programable automated fumigation cycles for use with both formaldehyde fumigation systems and hydrogen peroxide, using a Steris hydrogen peroxide generator. This will allow for remote venting of the cabinet and have evidence of the success/failure of cycles, including the cycle times. The fumigation sockets must remain live throughout the cycle.<br/>• The cabinet will be used as an aerosol test chamber and the control system will allow for modifications to the normal operation of the cabinet, this will include allowing for clean work to be undertaken without the fans in operation and independent operation of the fitted dampers.

Timeline

Publish date

11 months ago

Buyer information

UK Health Security Agency

Email:
science.tendering@ukhsa.gov.uk

Explore contracts and tenders relating to UK Health Security Agency

Go to buyer profile
To save this opportunity, sign up to Stotles for free.
Save in app
  • Looking glass on top of a file iconTender tracking

    Access a feed of government opportunities tailored to you, in one view. Receive email alerts and integrate with your CRM to stay up-to-date.

  • ID card iconProactive prospecting

    Get ahead of competitors by reaching out to key decision-makers within buying organisations directly.

  • Open folder icon360° account briefings

    Create in-depth briefings on buyer organisations based on their historical & upcoming procurement activity.

  • Teamwork iconCollaboration tools

    Streamline sales workflows with team collaboration and communication features, and integrate with your favourite sales tools.

Stop chasing tenders, start getting ahead.

Create your free feed

Explore similar tenders and contracts

Browse open tenders, recent contract awards and upcoming contract expiries that match similar CPV codes.

Explore other contracts published by UK Health Security Agency

Explore more open tenders, recent contract awards and upcoming contract expiries published by UK Health Security Agency.

Explore more suppliers to UK Health Security Agency

Sign up